Refrigerator, mechanical, blood bank
TXT
وصف:
Biologic Mechanical Refrigerator
MRC:
An overall measurement between specified points of an item,in distinction from height.ABFY
Overall Depth:
25.000 inches
MRC:
The distance measured in a straight line from the bottom to the top of an item.ABKW
Overall Height:
36.000 inches
MRC:
An overall measurement taken at right angles to the length of an item,in distinction from thickness.ABMK
Overall Width:
24.000 inches
MRC:
Indicates the type of current whether alternating,direct,or both.ACDC
Current Type:
Ac
MRC:
The value(s) of potential for which the item is rated.AMSE
Voltage Rating:
Between 115.0 volts and 230.0 volts
MRC:
The element,compound,or mixture of which the outside shell is fabricated,excluding any surface treatment.AYSX
Outside Shell Material:
Steel, corrosion resisting
MRC:
A measurement of the capacity of an item and its location.BBDY
Capacity And Location:
5.400 cubic feet refrigerated storage
MRC:
Those unusual or unique characteristics or qualities of an item not covered in the other requirements and which are determined to be essential for identification.FEAT
Special Features:
Shall contain a of 1/5 horsepower condensing unit; shall be capable of maintaining a preset temp range of from 2 degrees to 4 degrees (when fully loaded) ; shall come equipped with an audible and visual temp monitor; monitor shall be either ac or dc operated
MRC:
Describes the capabilities,intended use,and/or purpose for which the item is provided.FTLD
Functional Description:
Biological refrigerator desinged for use to contain laboratory reagents or human blood; shall be capable of holding and maintaining , at a constant temperature .60 450ml blood bags
MRC:
Characteristics or qualities of an item,not covered in any other requirement,which are considered essential information for one or more functions excluding nsn assignment.SUPP
Supplementary Features:
Shall give an instantaneous audible and visual warning when the temp within item falls below 1 degrees c or rises above 6 degrees c; shall include a fallsafe switch which silences the audible alarm and signals when safe temp is restored